Thermo Fisher Applied Biosystems Genetic Analyzers
Thermo Fisher Applied Biosystems genetic analyzer data collection software contains a file integrity vulnerability. An attacker with local access could modify .fsa/.hid output files containing DNA sequence data without detection. The affected products lack digital signature verification on output files, allowing modification of test results. Thermo Fisher has released security updates that implement digital signatures to verify file integrity and prevent tampering. Multiple data collection software versions for the 3500/3500xL, 3730/3730xL, SeqStudio Genetic Analyzer, SeqStudio Flex, and GeneMapper ID-X products are affected.
- Local access to the genetic analyzer data collection computer
- Write access to the output file directories on the instrument or connected computer
Patching may require device reboot — plan for process interruption
/api/v1/advisories/b06efe98-9d3c-465a-bf0e-113887582ac7Get OT security insights every Tuesday
Advisory breakdowns, a weekly summary, and incident analyses for the people actually defending OT environments. Free, no account required.